General procedures
Vivid E80/E90/E95 – Service Manual 4-7
GC091052 Rev. 3
Complete power down
1. Before Power Down, lock the Top Console in its lower,
locked position. This is required if you are going to move or
transport the Vivid E80/E90/E95 to ensure maximum
stability.
NOTE: For service purposes, you may want to move the Top
Console after Power Down. If you leave the Top Console in
unlocked position, it will be sensed during Power Down, and
the brakes that usually stabilize the position of the Top
Console, will be released so you can move the Top Console
in the XY directions.
2. Press once on the On/Off key on the Operating Panel to
display the System - Exit menu.
Figure 4-2. Press once on the On/Off key
Figure 4-3. Select Shutdown