Back End Processor (BEP) parts replacement
Vivid E80/E90/E95 – Service Manual 8-97
GC091052 Rev. 3
Back End Processor (BEP) replacement
Warnings
WARNING
Energy Control and Power Lockout for Vivid E80/E90/E95.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.
WARNING
When you return the used BEP to your local parts organization,
make sure you remove ALL PATIENT DATA from the hard
drive, given that the hard drive is still functional.In some
countries, you may be required to delete all software from the
disk before returning the BEP to the parts warehouse. Follow
your local policies.