Replacement Procedures
8-34 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Verification
Perform the following steps to verify that the product is
functioning as intended after this LCD Monitor replacement:
1. Verify that all screws that you removed earlier have been
installed.
2. If finished, connect the cables and probes you removed
earlier.
3. Power up the Ultrasound system to verify that it operates as
intended.
4. Move the LCD Monitor Arm from side to side and ensure
that it moves as intended.