Replacement Procedures
8-62 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
4. Remove the Rotary buttons below the Touch screen and the
Volume button.
5. Remove the Operating Panel, Upper.
6. Remove the Operating Panel, Lower. Place it on a clean
surface with the front down.
7. Remove Handle Left Top and Handle Right Top.
8. Remove the Palm Rest Cover.
Remove the Button IF Board Assy
Follow this procedure to remove the Button IF Board Assy
1. Disconnect all the five (5x) connectors from the board.
2. Push the TABs away to release the board.
3. Remove the board.
Install the Button IF Board Assy
Follow this procedure to install the Button IF Board Assy
1. Install the Button IF Board into the UI Frame, Lower by
pressing it under the small hooks.
2. Ensure the XYZ Buttons Frame cable runs along the Lower
Op Panel cable channel so that the cable is not pinched
when the Palm Rest is replaced.
3. Install the Handle Left Top and Handle Right Top.
4. Replace the Palm Rest.
5. Replace the Lower Operating Panel.
6. Replace the Upper OP Panel/Touch Panel assembly.

Table of Contents

Other manuals for GE Vivid E90

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E90 and is the answer not in the manual?

GE Vivid E90 Specifications

General IconGeneral
BrandGE
ModelVivid E90
CategoryMedical Equipment
LanguageEnglish

Related product manuals