EasyManuals Logo

GE Vivid E90 User Manual

GE Vivid E90
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #325 background imageLoading...
Page #325 background image
Back End Processor (BEP) parts replacement
Vivid E80/E90/E95 – Service Manual 8-103
GC091052 Rev. 3
Preparations (continued)
When preparing for the replacement, you must perform the
following steps:
1. Record the Ultrasound system’s TCPIP settings and
installed Option strings.
2. Back up the Patient Archive images, Report Archive,
User-Defined Configuration (Customer Presets), and
Service settings. You will perform a Restore after the install.
3. Reload the system software from removable media. Select
No when asked: Keep User Data and Patient Archive?
4. Power down the Ultrasound system.
5. Disconnect the Mains Power Cable from the wall outlet.
6. Disconnect all external cables (probes, ECG, physio,
Ethernet, audio, video, USB).
7. Remove the Left Side Cover.
8. Remove the BEP Cover.
Hard Disk Drive (HDD) removal procedure
Follow these steps to remove the HDD:
1. Disconnect the cables connecting the HDD.
2. Remove the two screws that secures the HDD Bracket to
the BEP’s frame.
3. Get hold of the HDD assembly and pull it out.

Table of Contents

Other manuals for GE Vivid E90

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E90 and is the answer not in the manual?

GE Vivid E90 Specifications

General IconGeneral
BrandGE
ModelVivid E90
CategoryMedical Equipment
LanguageEnglish

Related product manuals