Cleaning and disinfection
Vscan AirUser Manual 6-17
Direction GP092020-1EN Rev 18
GEHC ultrasound transducers web site
The GEHC Ultrasound transducers web site contains a list of
chemicals that have been tested for compatibility with GEHC
ultrasound probes. Verify probe compatibility via the GEHC
ultrasound transducers web site. The reprocessing instructions
provided in this document have been validated with the
chemicals specified in ‘Chemicals used for Efficacy Validation
with Vscan Air CL’ on page 6-7.
GEHC ultrasound transducers web site
https://gehealthcare.com/probecare
OR
https://www.gehealthcare.com/products/ultrasound/ultrasound-transducers
WARNING
Creutzfeldt-Jakob disease
This device is not indicated for neurological use. Neurological
contact on patients with this disease MUST BE avoided. If the
Vscan Air CL becomes contaminated, there is no adequate
means to disinfect it. In this case, the contaminated device/
probe MUST BE discarded in accordance with local biological
waste hazard procedures.