EasyManua.ls Logo

Gemmy PLC-01 - Speed and RCF Reference Tables

Gemmy PLC-01
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Speed and RCF Indication
The speed and RCF may be varied centrifuge to centrifuge caused by
its surrounding power consumption, which will result in a fluctuation of
speed. To assure an accurate rpm and its corresponding RCF, do not
place this equipment in the draft, sunlight or near a piece of equipment
that emits heat as well as Electromagnetic conduction emission.
1. Motor 67W (Average values from 5 random samplings)
2. Motor 90 W (Average values from 5 random samplings)
- Confidence interval: ± 95%
- User is recommended to measure speed and RCF by digital speed
indicator to achieve the practical values.
8
TIMER
PLC-01 ~ PLC-03
PLC-04~PLC05
Speed
RCF
Speed
RCF
LOW
1026
124 x g
1018
132 x g
1
1130
150
1123
160
2
1242
182
1262
202
3
1355
216
1603
326
4
1553
285
1810
416
5
1855
406
2055
536
6
2350
651
2280
660
7
3024
1079
2760
967
8
3447
1401
3100
1220
9
4030
1916
3585
1631
HIGH
4460
2346
3960
1990
TIMER
PLC-01 ~ PLC-03
PLC-04 ~ PLC-05
Speed
RCF
Speed
RCF
LOW
1000
118 x g
1000
127 x g
1
1056
132
1080
148
2
1165
160
1170
174
3
1305
201
1270
205
4
1485
260
1390
245
5
1672
330
1543
302
6
1940
445
1770
398
7
2405
682
2016
516
8
2863
967
2628
876
9
3815
1717
3550
1600
HIGH
4500
2390
4003
2033

Related product manuals