Troubleshooting:
Problem
Possible cause
Vibration
- Rotor imbalance
- Defective rubber support
Erratic speed
- Worn out carbon brushes
- Defective triac or VR
Failure of
operation
- Worn out brushes
- Defective motor or PCB
- Blowout fuse
- Disengaged linkage
Burnout motor &
PCB
- Wrong power supply
- Overload current
Change of MAX
and MIN rpm
- Loose SVR of PCB
Timer can’t work
- Counting failure
Tilted rotor
- Loose fastener attached to
motor
- Broken rubber support
Failure of motor
spinning
- Worn out carbon brushes
- Malfunction of PCB
- Burnout motor
- Disengaged wiring
NOTE
Troubleshooting must be done by the qualified personnel only.
The drive motor and linkage assembly do not require lubrication
Immerse the centrifuge in water is strictly prohibited
Disassembly must be performed under the supervision of licensed
engineers
Tip the equipment up or place upside down or vibrate the
equipment may damage the equipment
Take care of the electrical shock for troubleshooting.
Always refer troubleshooting to your local distributors.
Unstable voltage and current can shorten the lifespan of this
centrifuge. We recommend you use the voltage stabilizer.