Safety information
Your GEMMY Table Top Centrifuge has been designed with functions,
reliability, and safety in mind. This manual contains important operation
and safety information. The users must carefully read and understand
the contents of this manual prior to installing your centrifuge in
conformance with your local electrical codes.
Classification: This Centrifuge is classified as “Class I” equipment.
Thermal cutout operation temperature
105ºC
CE marking
The manufacturer also declares the conformity with the
actual required safety standards in accordance with
LVD 73/23 EEC.
Standard:
This equipment complies with EC Electromagnetic Compatibility
requirement of EN60601-1-2. EMI filter is designed in conformity
with 89/336 EEC-EMC Directive.
ISO9001+13485,
GMP (Good Manufacturing Practice)
FDA registered (USA)
Product’s briefing
This equipment is a Class 1 tabletop centrifuge with timers of 60
minutes maximum and a speed control knob. The motor has a max
speed of 4000 4500 rpm and fitted with a thermal cut-off to cut the
power in abnormal temperature condition. There is no heating
element. The power cord is non-detachable. Two PC boards are
installed, one is for power filtering, one is for the speed control. There
is no applied part for patients and not suitable for the use in special
environment.
Motor
AC, carbon brush, your choice of 67W or 90W
power
110V/60HZ or 220V/50HZ
rotor
45º fixed angle rotor
Suggested Tubes:
5ml: 12 x 75mm
10ml: 16 x 100mm Length
15ml: 17 x 117mm (or120mm)