EasyManua.ls Logo

Gemmy PLC-01 - Speed Recalibration Procedure

Gemmy PLC-01
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Recalibration Procedure for Speed:
This centrifuge was initially set by its speed from 1000 4500 rpm (or
4000 rpm for PLC-05) in house. Supposed that the min/max speed is
varied, the control can be recalibrated as depicted below:
1
Remove the bottom plate, switch power on by turning control knob
of timer. CAUTION: Electrical Shock
2
Turn knob of speed regulator to the extreme LOW position. Adjust
part of SVR P2/204 (for the low 1000 rpm control) right/left to get
the desired speed of 1000rpm.
3
To recalibrate the speed for 4000 or 4500rpm, turn speed regulator
to the extreme HIGH position, then adjust SVR (P1/104) right/left
slightly until it reaches 4000 or 4500 rpm.
4
The calibration of LOW or HIGH speed can influence each other;
thus the user is recommended to use the digital speed indicator
and measure the result of the recalibration. Repeat the above
procedure until the final speed indication meets your requirement.
5
Switch off the power and assemble the bottom plate.
6
Glue the parts of SVR (P1/104 and P2/204) after the speed
recalibration is finished. The transportation and vibration may
probably loosen the electronic parts and fasteners.
WARNING:
Electric shock and Personal injury. Always proceed with caution.
7

Related product manuals