GETINGE-LA CALHENE ISOTEST | User manual
NTA
3015/78 C
Approuvé informatiquement / Electronic signature
DEX | Page 40/
74
Step 12: The test phase is correct.
If the leak test phase is correct, the biodecontamination cycle starts and the
following animation appears:
Step 13: If the biodecontamination phase is completed with no errors, the airlock enters
the unloading phase.
The airlock is now sterile.
The following animation appears: