Page #40 background imageLoading...
Page #40 background image
5
Recording the inspection
SW Service Protocol PM OR Lights LUCEA 50 100-A
40 / 44
Lucea 50_100
01740 EN 01
4
www.getinge.com
Form:
STI-P01
-F11
-B-en
Pass
Fail
N/A
Continuity test 1
Continuity test 2
Pass
Fail
N/A
All LEDs operate correctly
ON / OFF
(Light head button)
Illumination increase / decrease level
(Light head capacitive keypad)
Wireless remote control operates correctly
Camera operates correctly
Switch over to battery mode and back to mains
Batteries charge indicator brights when light is off
Illumination
Acceptable
values (klx)
Measured value
(klx)
Pass
Fail
N/A
Ec Max LUCEA 50
40 < Ec < 160
Ec Max LUCEA 50
40 < Ec < 160
Ec Max LUCEA 100
70 < Ec < 160
Ec Max LUCEA 100
70 < Ec < 160
Limit ()
Measured values ()
Pass
Fail
N/A
Protective earth continuity 1
300
Protective earth continuity 2
300