5
Recording the inspection
SW Service Protocol PM OR Lights LUCEA 50 100-A
40 / 44
Lucea 50_100
01740 EN 01
Document name: SW Service Protocol PM OR Lights LUCEA 50 100
8. Optical evaluation
Measure the illumination at the center, at 1 meter, small light patch (with dimmer at max setting).
The IEC 60601-2-41 standard indicates the limits between 40 000 lux minimum & 160 000 lux maximum.
The minimum acceptable value has been calculated from the nominal illumination values -30%.
9. Electrical Safety Tests (IEC 62353)
Test records should be enclosed to this report for any further use, when available.
10. Function test
All LEDs operate correctly
ON / OFF
(Light head button)
Illumination increase / decrease level
(Light head capacitive keypad)
Wireless remote control operates correctly
Camera operates correctly
Switch over to battery mode and back to mains
Batteries charge indicator brights when light is off
Protective earth continuity 1
Protective earth continuity 2