GETINGE SAFESTEP ATP MONITOR GETINGE SAFESTEP ATP MONITOR
27 28
ATP Adenosine Triphosphate – energy carrier
molecule
Device The Test® Swab sample collection and
chemistry reaction device
EMC Electro-Magnetic Compatibility
fmol Femtomole (10-15 moles)
HACCP Hazard Analysis Critical Control Point
LCD Liquid Crystal Display
NiCD Nickel Cadmium rechargeable baeries
NiMH Nickel Metal Hydride rechargeable baeries
PC IBM compatible personal computer of
Pentium 1 specification or higher, running
Windows 98 or later
Reading Measurement value in RLUs
Result Measurement pass (), fail () or caution ( ! )
RLU Relative Light Units (unit of measurement)
RS232 Serial communications protocol for
connecting the unit to a PC, used for
uploading test results data
USB Universal Serial Bus used to connect
computer peripherals to a PC
Unit The Getinge SafeStep ATP Monitor unit
General:
Unit dimensions (W x H x D) 72mm x 191mm x 32mm
Unit weight (including baeries) approx. 260g
Operating temperature range
Relative Humidity range
5°C to 40°C
20 - 85%, non-condensing
Storage temperature range
Relative Humidity Range
-10°C to 40°C
20 - 95%, non-condensing
Unit Details:
Measurement range 0 to 9999 RLUs
Measurement resolution 1 RLU
Measurement time 15 seconds
Measurement noise ±5% or ±5 RLUs
Programmable result thresholds 251 programs
Programmable location names 251 names of 20 characters
Results memory size 2000 tests
Serial interface EIA-232 compatible
Baeries:
Baery Size (2 off) AA, LR6 or E91
Baery Types
Non-rechargeable
Rechargeable
(externally charged)
nom. 1.5V Alkaline
nom. 1.2V NiMH or NiCD
Baery Capacity (for 2600mAh)
Standby mode (at 20°C)
Continuous reading
min. 6 months
min. 500 tests
9. Glossary of Terms and
Abbreviations
10. Technical Specifications