GETINGE SAFESTEP ATP MONITOR GETINGE SAFESTEP ATP MONITOR
9 10
TIP: The unit can be re-zeroed by removing the Test® Swab
device, closing the lid and then pressing and holding down
the “OK” buon for 1 second. The unit will then perform a
15-second zero measurement.
3.3 Performing a Retest
When viewing a failed ( ) test result (see section 4.5), the
program location can be selected for retesting by pressing
the “P” buon, followed by the “OK” buon:
NOTE: It is only possible to select a program location for
retest if the most recent test result for that location was a
failure ( ).
Having selected retest mode (signified by the
icon next
to the PROG#), perform the sample measurement in the
usual way, using a new Test Swab device.
Once the retest measurement is complete, the original
failed test result is marked with the
icon, and the new
test result is tagged with the icon:
NOTE: Retest mode is automatically deselected once
the new measurement is complete, or can be manually
cancelled by pressing the “P” buon and selecting a new
program location (see section 4.2).
3.4 Standby Mode
The Standby Mode is used to control when the unit should
automatically turn off when not in use. This is designed to
conserve baery power.
The default is 20 minutes. If no buons are pressed
within the time period, the unit will turn off. If any buon
is pressed, or a reading is performed, the Standby Mode
countdown timer is reset to zero.
If the bell icon is enabled the unit will begin beeping every
10 seconds for 1 minute as a warning before turning off. For
example, if the Standby Mode time is set to 20, then if the
unit is not used for 19 minutes it will start to beep every 10
seconds for 1 minute. If no action is taken the unit will turn
off. If any buon except the power buon is pressed the
countdown will restart.
To change the Standby Mode seing, select the STANDBY
MODE menu option by pressing the
buon. Scroll to
“STANDBY MODE” and press “OK” buon. Then use the up
or down buon to increase or decrease the Standby Mode
duration, followed by the “OK” buon to store the new
seing. Standby Mode duration can be from 5 to 60 minutes
in increments of 5 minutes. Then use the up or down buon
to enable or disable the warning beeps. This will toggle the
bell icon from enabled
to disabled . Press the “OK”
buon to accept the seings:
increase or decrease
duration (5 to 60)
set duration
enables or disable
warning beeps
set warning state
3.5 Bold Text
The unit displays key information on the LCD is bold to
improve legibility. In some cases the number and size of the
text will exceed the alloed space on the LCD. If bold is set
to OFF all text on the unit will be displayed normally. This
will allow all text to be displayed even if the text is maximum
and all wide characters.
To change the Bold Text seing, select the BOLD TEXT
menu option by pressing the
buon. Scroll to “BOLD
TEXT” and press “OK” buon to switch between ON and
OFF:
OFF
ON
EXIT
3.6 Save Tests
By default any test performed on the unit is recorded to the
unit’s non-volatile memory. Even if you take the baeries
out the results of the test are retained in the unit’s memory.
In some cases, such as training, calibration checks, etc.
users may want to discard test results. If Save Test is set to
OFF, all tests taken in this mode will not be recorded.
NOTE: The Save Tests mode will be set back to ON when
the unit is turned off and back on again. When Save Tests
mode is set to OFF “RESULTS NOT SAVED” will appear at the
boom of the LCD.
To change the Save Tests seing, select the SAVE TESTS
menu option by pressing the up buon. Scroll to “SAVE
TESTS” and press “OK” buon to switch between ON and
OFF:
OFF
ON
EXIT