8.3 Bluetooth
®
Interface (only with feature M01)
The Bluetooth
®
interface facilitates various functions at the test instrument:
• Data transmission (test structure and measurement data) to IZYTRONIQ test software (see section 19)
• Transfer of results to IZYTRONIQ test software (push-print function) (see section 15.4 for single measurements and section 17.7 for
test sequences)
Bluetooth
®
settings are entered under SETUP 3/3 > Bluetooth:
Security
For security reasons, we recommend deactivating the
Bluetooth
®
interface if it’s not needed. (The “not visible”
setting cannot be used as a substitute for shutting down
the Bluetooth
®
interface, because invisible Bluetooth
®
devices can nevertheless be found using the appropriate
means.)
Furthermore, device pairings which will no longer be
required for a lengthy period of time should be deleted.
Use of Multiple Test Instruments (risk of mix-up)
The test instrument’s device name is set to “SECUTEST”
as a default setting. If you access one PC with several
test instruments, their names should differ (or at least be
individually supplemented: SECUTEST1, SECUTEST2
etc.).
Menu selection for operating parameters, page 3 of 3
Setup 3/3
Menus for using the Bluetooth
®
interface
Menu selection for Bluetooth
®
operating parameters
Specifies whether the test instrument can be
found by other Bluetooth
®
devices.*
The name of the test instrument displayed via
the interface can be changed here. *
Find/pair Bluetooth
®
devices, view/edit existing
pairings *
* These submenus only appear if status is set to on.
Search for nearby Bluetooth
®
devices
Paired device found
(white frame) > rename or delete
Not yet paired device found
(blue frame) > pairing PIN entry
Paired device found
(blue frame) > rename or delete
List of already paired devices
Switch Bluetooth
®
interface on/off
Softkeys are keys with functions that vary depending
upon the menu level.
Function Keys
PRINT
ESC
HELP
MEM
PRINT
ESC
HELP
MEM
PRINT
ESC
HELP
MEM