68 Gossen Metrawatt GmbH
15.9 Probe Voltage – U
Mains to Test Socket
Schematic Diagram
Wiring Diagram
Permanently Connected DUT
Schematic Diagram
Wiring Diagram
Direct, alternating and pulsating voltages of up to 253 V can be
measured. Two connection types are available, one of which has
to be selected in the parameters menu.
Setting Measuring Parameters for U
Probe
Test Procedure
Set the rotary switch to the U position.
Connect the DUT’s mains plug to the test instrument’s test
socket.
Start the test: Press the START/STOP key.
PE-P1 (with mains): Acknowledge the warning which in-
dicates that line voltage will be connected to the test
socket.
Switch the device under test on.
Contact the ungrounded output for safety extra-low voltage
with test probe P1.
Polarity can be set via direct selection immediately be-
fore measurement is started, without having to switch to
the parameters menu.
The measured values are displayed. The measured
value recording icon shown at the right appears. Each
time this key is pressed, the currently displayed mea-
sured value is saved to buffer memory.
Turn off the device under test.
End the test: Press the START/STOP key.
The save icon appears (floppy disk showing the num-
ber of measured values stored to buffer memory) and
prompts you to save the measured values to a test ob-
ject ID number.
Press the ESC key in order to discard the measured
values stored to buffer memory and acknowledge by
pressing the key shown at the right.
Single Measurements, Rotary Switch Level: Green
Rotary Switch
Position
Meas. Type
With Mains to
Test Socket
Meas. Type
Without
Mains to Test
Socket
Measuring Functions
U
PE - P1
UProbe voltage, RMS
U
~
Alternating voltage component
U
=
Direct voltage component
PE - P1
(with mains)
UProbe voltage, RMS
U
~
Alternating voltage component
U
=
Direct voltage component
U
Measuring
Parameter
Meaning
Meas. Type
Suitable for
DUT Connection via
PE-P1 Measurement of voltages with
reference to PE, test socket re-
mains voltage-free
Permanent connection
PE-P1 (with mains) Measurement of voltages with
reference to PE, line voltage is
applied to the test socket
Test socket
Polarity – only for PE-P1 (with mains)
Normal / reversed Selection of polarity for mains voltage to the test socket