Gossen Metrawatt GmbH 55
Wiring Diagram
Alternative Measuring Method (equivalent leakage current)
Alternative measurement type
– DUT mains plug (protection class I) to test socket
Schematic Diagram
After activating test voltage, leakage current is measured via the
DUT’s mains cable between short-circuited mains conductors L
and N and the protective conductor terminal at the DUT.
Wiring Diagram
Connection of 3-Phase DUTs (only with feature I0, e.g. SECUTEST
ST PRO, and with optional AT3-IIIE test adapter)
AT3-Adapter measurement type
– DUT mains plug to AT3-IIIE test adapter
– AT3-IIIE probe to COM-V terminals
– AT3-IIIE test plug to test socket
Schematic Diagram
Measurement of the DUT with 3-phase mains connection via
AT3-IIIE adapter
Wiring Diagram (AT3-IIIE probe to COM-V)
Measurement of Protective Conductor Current via Current Clamp
Sensor with Voltage Output for Permanently Installed DUTs
(only with feature I0, e.g. SECUTEST ST PRO, and with optional
current clamp sensor)
Clamp measurement type
Schematic Diagram
Measurement of protective conductor current by closing the cur-
rent clamp sensor around mains cable PE for permanently
installed protection class I DUTs
Wiring Diagram (current clamp sensor to COM-V)
Set Measuring Range at Clamp Meter and Parameter at SECUTEST ST
PRO
*Only with WZ12C
** Default value
SECUTEST ST PRO Clamp SECUTEST ST PRO
Transformation
ratio parameter
Transformation
ratio
(switch *)
Measuring range Display range
with clamp
1 mV : 1 mA
WZ12C
1 mV : 1 mA 1 mA ... 15 A 0 mA ... 300 A
100 mV : 1 mA
SECUTEST CLIP
100 mV : 1 mA 0.1 ... 25 mA 0.00 mA ... 3.00 A