82 Gossen Metrawatt GmbH
Classification Parameters – VDE 0701-0702-VLTG /
ÖVE E 8701-VLTG / SNR 462638-VLTG
1
These parameters must be entered manually if they’re not automatically de-
tected, or if they’re detected incorrectly.
2
The limit value for protective conductor resistance is determined on the ba-
sis of length and cross-section (length only in the case of EL1). Data remain
in memory until a new entry is made.
Classification Parameters – VDE 0701-0702-PRCD /
ÖVE E 8701-PRCD / SNR 462638-PRCD
1
These parameters must be entered manually if they’re not automatically de-
tected, or if they’re detected incorrectly.
2
The limit value for protective conductor resistance is determined on the ba-
sis of length and cross-section. Data remain in memory until a new entry is
made.
3
New classification parameters “PRCD type”
(only displayed if “VDE 0701-0702-PRCD / ÖVE E 8701-PRCD /
SNR 462638-PRCD” standards parameter has been selected):
PRCD (standard): For the testing of simple circuit breaker safety adapters in
which the protective conductor is permanently connected. Commonly desig-
nated 2-pole.
PRCD (SPE): (SPE = switched protective earth) for testing PRCDs in which the
protective conductor is only connected in switched-on condition. Commonly
designated 3-pole.
PRCD-S (SPE): For testing type PRCD-S circuit breaker safety adapters.
PRCD-K (SPE):
For testing type PRCD-K circuit breaker safety adapters.
For more information on testing single-phase and 3-
phase type S and K PRCDs by simulating faults see
“PROFITEST PRCD test adapter” on our website.
Testing of PRCDs (test procedure and time to trip) is only
possible for DUTs with a nominal voltage of 230 V.
Classification Parameters – NEN 3140
(only with feature D01/D02 or corresponding enabling for a
fee)
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
The limit value for protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new entry
is made.
Parameter Setting Options / Meaning
1/2
Standard / Test Sequence Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences”, for a list of standards available
for selection in the test instrument)
Protection class
1, 2
PCI
Connection type
1, 2
Test socket
Adapter: AT3-IIIE
Adapter: EL1 adapter
Adapter: VL2E adapter
2/2
Measurement type (MT)
1
VLTG
2
Auto-detection Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & meas. type
Protection class only (PC)
Measurement type only (MT)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)
Parameter Setting Options / Meaning
1/2
Standard / Test Sequence Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences”, for a list of standards available
for selection in the test instrument)
Protection class
1, 2
PCI, PCI+II
Connection type
1, 2
Test socket
2/2
Measurement type (MT)
1
PRCD
3
PRCD type
3
PRCD (standard)
PRCD (SPE)
PRCD-S (SPE)
PRCD-K (SPE)
Auto-detection Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & meas. type
Protection class only (PC)
Measurement type only (MT)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)
Parameter Setting Options / Meaning
1/2
Standard / Test Sequence Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences”, for a list of standards available
for selection in the test instrument)
Protection class
1, 2
PCI, PCII, PCI+II, PCI+III, PCII+III, PCI+II+III
Connection type
1, 2
Test socket
Permanent connection
Adapter: AT16/32-DI adapter
Adapter: VL2E
Adapter: AT3-Adapter (feature I01)
Permanent connection: P1+P2 (only with feature H01)
2/2
Measurement type (MT)
1
Passive
Active
Auto-detection Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & meas. type
Protection class only (PC)
Measurement type only (MT)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)