86 Gossen Metrawatt GmbH
Note
For more information on testing single-phase and 3-
phase type S and K PRCDs by simulating faults see
PROFITEST PRCD test adapter” on our website.
Note
Testing of PRCDs (test procedure and time to trip) is only
possible for DUTs with a nominal voltage of 230 V.
Classification Parameters – IEC 62368/ EN 62368 / VDE 0868-1
(Only with enabling of Z853U, “EN50678/50699 update”
or with feature KE01)
1
These parameters must be entered manually if they’re not automatically de-
tected, or if they’re detected incorrectly.
Classification Parameters – IEC 62911/ EN 62911 / VDE 0868-911
(Only with enabling of Z853U, “EN50678/50699 update”
or with feature KE01)
1
These parameters must be entered manually if they’re not automatically de-
tected, or if they’re detected incorrectly.
Classification Parameters – IEC 60974-4/
EN 60974-4 / VDE 0544-4
1
These parameters must be entered manually if they’re not automatically de-
tected, or if they’re detected incorrectly.
Classification Parameters –
IEC 62353 / EN 62353 / VDE 0751-1
1
These parameters must be entered manually if they’re not automatically de-
tected, or if they’re detected incorrectly.
Parameter Setting Options / Meaning
1/2
Standard / Test Sequence Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences”, for a list of standards available
for selection in the test instrument)
Protection class
1
PCI, PCII, PCI+II
Connection type
1
Test socket
Permanent connection
Adapter: AT16/32-DI adapter
2/2
Measurement type (MT)
1
Active
Auto-detection Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & meas. type
Protection class only (PC)
Measurement type only (MT)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)
Parameter Setting Options / Meaning
1/2
Standard / Test Sequence Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences” for a list of standards available
for selection in the test instrument)
Protection class
1
PCI, PCII, PCI+II
Connection type
1
Test socket
Permanent connection
Adapter: AT16/32-DI adapter
2/2
Measurement type (MT)
1
Active
Auto-detection Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & meas. type
Protection class only (PC)
Measurement type only (MT)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)
Parameter Setting Options / Meaning
1/2
Standard / Test Sequence Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences”, for a list of standards available
for selection in the test instrument)
Protection class
1
PCI, PCII, PCI+II
Connection type
1
Test socket
Permanent connection
Adapter: AT16/32-DI adapter
Adapter: AT3-Adapter
2/2
Measurement type (MT)
1
Active
Voltage rating,
serial plate
U(R)eff (= serial plate rating)
U(0) DC (open circuit voltage, DC, fixed: 113 V)
Auto-detection
Conn. & PC & MT (not recommended as MT selection is
fixed)
Connection & PC
Conn. & MT (not recommended as MT selection is fixed)
Connection only
PC & MT (not recommended as MT selection is fixed)
Protection class only (PC)
Meas. type only (MT) (not recommended as MT selection
is fixed)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)
Parameter Setting Options / Meaning
1/2
Standard / Test Sequence Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences”, for a list of standards available
for selection in the test instrument)
Protection class
1
PCI, PCII or PCI+II
Connection type
1
Test socket
Permanent connection
Adapter: AT16/32-DI adap.
Adapter: AT3-Adapter
Permanent connection: P1+P2 (only with feature H01)
2/2
Measurement type (MT)
1
Passive active
APPs Applied parts: none, B, BF, CF or combinations
Type B (body): DUTs of this type are suitable for both in-
ternal and external patient applications, except for use in
direct proximity to the heart.
The following protection classes are permissible:
I, II, III or devices with internal electric power supply.
Type BF (body float): Same as type B, but with type F in-
sulated applied parts
Type CF (cardiac float): DUTs of this type are suitable for
use directly at the heart. The insulated applied part may
ungrounded.
The following protection classes are permissible:
I, II or devices with internal electric power supply.
Auto-detection Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & meas. type
Protection class only (PC)
Measurement type only (MT)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)