CWT7 USER AND SERVICE MANUAL
Revision: 03/2018 Page: 13
1 1 0022 (both kg and lb units are enabled with calibration using unit lb, dual range)
1 1 0 122 (In addition, PCS is on, and therefore used as a counting scale)
Press【Print】key to set the capacity of the scale 0 00 3 0 0
Press【Print】key to set the decimal point d P 0. 00
Using【Zero】key or【Tare】key to shift it from 0.0 until 0.00000
Press【Print】key to set the division d i v 0 1
Using【N/G】key to shift it between 01/02/05/10/20/50
LF3 Linearity Calibration
Press【Print】key to set the linearity calibration W 0
Press
【
ON/T
】
key to enter next step W 1
Put 1/3F.S. test weight and press
【
ON/T
】
to enter next step W 2
Put 2/3F.S. test weight and press
【
ON/T
】
to enter next step W 3
Put 100%F.S. test weight and press
【
ON/T
】
to enter next step W 4
Press 【Print】key to exit and back to LF- 3
LF4 A/D Converting Speed
The same operation as Weighing Speed
*It was blocked when UF-5 set of HOLD 1
*1=15Hz 2=30Hz 3=7.5Hz
LF5 Zero Tracking
The same operation as UF 8 Zero Track
*It was blocked when UF- 5 set of HOLD 1(animal weighing)
LF6 Type Approval
nonE is for non-certified scales, and oiML is for NTEP (United States) or OIML
(Europe)
LF7 Gravity Adjustment
The same operation as UF 9 Gravity Adjusting
LF8 Zero
Press【Print】key to set the initial zero function SEtZ Y
y Reset of the zero point each time when the scale switches on
n Disable resetting zero when switching on the scale