Possible Causes: Corrective Action:
METER READING DOES NOT TRACK FLOW
Sensor and GND wires reversed or not properly
connected
Check Sensor connections
Improper AC power input Ground
Use direct connection with 12 AWG wire to
nearest Ground pole (<1 ohm resistance).
Page 23