24 HR LOG (Data Logging option only)
Press Å from the RUN display to view a formatted flow report from
instruments with a built-in data logger. Press È to scroll down one day or
repeatedly to scroll to a specific date. Up to 365 days can be stored.
Newest date will overwrite the oldest. Press ü to return to the main
display.
PASSWORD
The password (a number from 0000 to 9999) prevents unauthorized access
to the Calibration menu.
From the Run display press the Æ key to get to
Password. Factory
default password is 0000 and if it has not been changed press the ü to
proceed to the Menu Selections screen.
If a password is required, press Æ to place the cursor under the first digit
and È or Ç to set the number, then Æ to the second digit, etc. Press Æ or
ü to proceed to the Menu Selections screen.
A new password can be stored by going to
Special Functions/New
Password
.
Page 9
DFM 5.0 Doppler Flow Meter
--Password----------
Password 0000
--24 hr log----------
>Date Feb. 12/2008
Total 0.000ft/s
Average 0.000ft/s
Maximum 0.000ft/s
Max Time 01:57:00
Minimum -1.046ft/s
Min Time 02:35:00