41
© Guldmann GB/US-12/2018 • # 550383_9.00
© Guldmann GB/US-12/2018 • # 550383_9.00
During the safety/service inspection a report must be prepared on what was
checked and replaced. Parts that are worn or defective must be replaced with
new parts from Guldmann. Spare parts drawings and documentation can be
obtained from the manufacturer or supplier.
Documentation/checklist regarding safety/service inspection can be obtained
from the manufacturer or supplier.
Re-verifying Class III scales
To maintain the medical approval of a Class III scale, the digital scale must
be re-veried/calibrated in accordance with local regulatory requirements and
by a Notied Body.
5.03 Troubleshooting
The GH3 does not respond to the hand control’s keys
1. Check the emergency stop is not activated
2. Check the hoist has power supply
3. Check the transformer is switched on and connected to the rail system
4. Contact the Guldmann Service Team
6.00 Classication
CE marking
Type B in accordance with UL/EN 60601-1
Read the manual before use
Must not be disposed of as standard household waste, must be
recycled.
Class I equipment: Permanent installation with protective ground
Class II equipment: Non-permanent installation without protective ground
The equipment is not suitable for use in the presence of ammable mixtures.
Degree of protection against harmfull ingress of liquids (water)
Hoist IP20
Hand control IP44
Remote control IP20
Transformer IP20