4.01 Where to use GL5.2
GL5.2 is designed to be used indoors whenever the user needs to be lifted or
moved on a level surface.
The lifter is designed so its legs can t under beds, around chairs / wheel-
chairs
The lifter can be used in wet areas. However, it must not be exposed to
splash water.
The lifter is designed to be used with multiple users. When moving the lifter to
another user / ward / room, perform normal hygienic disinfection.
4.02 Warning
Read the user manual fully before using the lifter to familiarize yourself with
the controls and safety features on the lifter.
Do not exceed the maximum load.
Use the lifter to lift a person only.
Use the lifter on an even and level surface only. The lifter has limited
mechanical stability when the lift is placed sideways on a slope.
Only use a lifting hanger approved for use with the lifter.
Don’t lift/turn the lifting hanger in vertical direction during the mounting of the
straps of the sling.
When adjusting the legs of the lifter make sure that no persons stand close to
the legs due to the risk of being jammed.
Do not run the lifter into persons or objects.
Exchange of lifting motor/actuator according to the manufacturer’s instructions.
See service chapter
In case of damage, do not use the lifter until authorized by qualied service
sta󰀨 or the Guldmann service team.
Do not use the lifter in areas where it can be splashed with water.
Do not use the lifter in the presence of a ammable anesthetic mixture with
air or with oxygen or nitrous oxide
Do not modify this equipment without the authorization from the manufacturer,
including opening battery which can damage cell and generate excessive
heat.
The lifter is not intended to drive over steps and thresholds.
The lifter needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in Chapter 15
EMC Information.
Portable and mobile RF communications equipment can a󰀨ect the lifter.
The use of ACCESSORIES, transducers and cables other than those speci-
ed, with the exception of transducers and cables sold by Guldmann A/S of
the lifter as replacement parts for internal components, may result in incre-
ased EMISSIONS or decreased IMMUNITY of the lifter.