GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
70
Buffer Size
The command receive buffer of instrument is 128 bytes. The return message buffer
of instrument is 1024 bytes.
If your combining commands size is larger than 128 bytes, commands will be lost.
If you query the waveform data (WA? or WB?), the instrument will return 1000
bytes waveform data. Do not use any query commands during the data transmission.
GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
71
Command List
Note: Every command must add 1 byte message terminator of LF (0x0A).
Function Transmission Format
*IDN? *IDN?
*CLR *CLR
*ULK *ULK
*RST *RST
Vertical Scale
V1=<Parameters (3bytes)><Target (1byte)>
V1? <Target (1byte)>
Input Coupling
V2=<Parameters (3bytes)><Target (1byte)>
V2? <Target (1byte)>
Display
V3=<Parameters (3bytes)><Target (1byte)>
V3? <Target (1byte)>
Probe
V4=<Parameters (3bytes)><Target (1byte)>
V4? <Target (1byte)>
V-VAR
V5=<Parameters (3bytes)><Target (1byte)>
V5? <Target (1byte)>
Invert
V6=<Parameters (3bytes)>
V6?
ADD
V7=<Parameters (3bytes)>
V7?