GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
⎯ ⎯
74
Details of Command Reference
Each command in this chapter will give a brief description. The examples of each
command will be provided and what query form might return.
Note: Every command must add 1 byte message terminator of LF (0x0A).
Common Commands
*IDN?
* IDN?
Queries the unique identification code of instrument
Return: GW,GRS60X2A,V.1.10
*CLR
* CLR
Clear the output buffer data of instrument
*ULK
* ULK
Unlock front panel control
GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
⎯ ⎯
75
*RST
* RST
Load factory default setting into instrument
Vertical Controls Commands
Vertical Scale
V 1 = Parameters (3Bytes) Target
V1?Target
Sets or queries the vertical gain of the specific channel.
Syntax
V1=010A
V1?B
Arguments
Parameters:
001→20V/DIV 002→10V/DIV 003→5V/DIV 004→2V/DIV
005→1V/DIV 006→.5V/DIV 007→.2V/DIV 008→.1V/DIV
009→50mV/DIV 010→20mV/DIV 011→10mV/DIV 012→5mV/DIV
013→2V/DIV 014→1mV/DIV
Target:
A→Channel 1 B→Channel 2