GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
78
Target:
AChannel 1 BChannel 2
V-VAR
V 5 = Parameters (3Bytes) Target
V5?Target
Sets or queries the vertical VAR function of the specified channel.
Syntax
V5=071A
V5=?B
Arguments
Parameters:
001 (uncalibrated, Max) to 71 (uncalibrated, Min) Enable vertical VAR function
and sets VAR value.
000 (calibrated) Disable vertical VAR function.
Target:
AChannel 1 BChannel 2
Invert
V 6 = Parameters (3Bytes)
GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
79
V6?
Sets or queries the invert function.
Syntax
V6=001
V6?
Arguments
Parameters
000Disable invert function
001Enable invert function
ADD
V 7 = Parameters (3Bytes)
V7?
Sets or queries the addition function.
Syntax
V7=001
V7?
Arguments
Parameters:
000Disable the addition function
001Enable the addition function