GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
⎯ ⎯
82
Arguments
Parameters:
If Target=A, Parameter range→0 (0 DIV) to 500 (-10DIV)
If Target=B, Parameter range→0 (0DIV) to 200 (4 DIV)
Target:
A→Pre B→Post
H-VAR
H 3 = Parameters (3Bytes)
H3?
Sets or queries the horizontal VAR function.
Syntax
H3=128
H3?
Arguments
Parameters:
080 (uncalibrated, Max) to 255 (uncalibrated, Min) →Enable horizontal VAR
function and sets VAR value
000 (calibrated) →Disable horizontal VAR function
GRS-6052A/6032A OSCILLOSCOPE
USER MANUAL
⎯ ⎯
83
XY Mode
H 4 = Parameters (3Bytes)
H4?
Sets or queries the XY mode
Syntax
H4=001
H4?
Arguments
Parameters:
000→Disable XY mode
001→Enable XY mode
Sweep Magnification
H 5 = Parameters (3Bytes)
H5?
Sets or queries the MAG function.
Syntax
H5=002
H5?