Command Reference
116
11.11.9 COMParator:OPEN
The :COMParator:OPEN command selects the open condition for main
parameter.
Command
Syntax
COMParator:OPEN {OFF,2,5,10,20,50}
Parameter
Where,
OFF Turns the beeper off.
2,5,10,20,50 The percent range value
Example
SEND> COMP:OPEN 2<NL>
Query Syntax
COMParator:OPEN?
Query Response
{OFF,2,5,10,20,50}
Example
SEND> COMP:OPEN?<NL>
RET> OFF<NL>
11.12 LIST Subsystem
The LIST or SWEEP Subsystem command group sets the List Sweep
measurement function, including the sweep point setting and limit values for
the limit function.
Figure 11-9 LIST Subsystem Command Tree
11.12.1 LIST:PARAmeter
The LIST:PARAmeter command sets the list sweep parameter.
Command
Syntax
LIST:PARAmeter {FREQ,VOLT,CURR}
Parameter
Where, {FREQ,LEVEL} is:
FREQ Sets the sweep parameter to frequency
VOLT Sets the sweep parameter to voltage level
CURR Sets the sweep parameter to current level
Example
SEND> LIST:PARA VOLT<NL>
Query Syntax
LIST:PARAmeter?
LIST
:PARAmeter
:STAT
{FREQ,VOLT,CURR}
<n>,{ON,OFF}
:BAND
<n>,<value>,{A,B,OFF},<low>,<high>