4.1.5 Test Signal Voltage/Current Level [LEVEL]
The LCR-6000 Series’ test signal voltage/current level
can be set as the effective value (RMS value) of a sine
wave of the test frequency from the unit’s internal
oscillator.
When a constant voltage or current level measurement is
performed, an asterisk mark (*) appears at the end of the
LEVEL display.
Test signal level setting procedure
Step 1. Press the [Measure] key
Step 2. Use the cursor key to select [LEVEL] field
Step 3. Use the soft-keys or numeric entry keys to
enter the test signal voltage/current level.
Soft-key
Function
INCR +
Refer to Table 4-13
DECR -
ALC ON
Auto Level Control ON
ALC OFF
Auto Level Control OFF
Table 4-13 Test voltage/current level that can be selected with
INCR+/DECR
INCR+/DECR-
10.00mV
100.0mV
300.0mV
500.0mV
1.00V
1.50V
2.00V
INCR+/DECR-
100.0uA
500.0uA
1.00mA
5.00mA
10.00mA
20.00mA
Voltage range
Resolution
10.00mV LEVEL 99.99mV
0.01mV
100.0mV LEVEL 999.9mV
0.1mV
1.00V LEVEL 2.00V
0.01V
Current range
Resolution
100.0uA LEVEL 999.9uA
0.1uA
1.00mA LEVEL 20.00mA
0.01mA