LCR-800 Series User Manual
52
F setting (FREQUENCY):
l Press F1/F2 key to select the white inverted characters “F”.
l Key-in testing frequency through panel keyboard.
l Press key shown as Figure 8-6 below. For further details, please refer to
4-5-2.
Figure 8-6
V setting (VOLTAGE)
l Press F1/F2 key to select the white inverted characters “V”.
l Key-in testing voltage through panel keyboard.
l Press key shown as Figure 8-7 below. For further details, please refer to
4-5-4.
Figure 8-7
EXIT
SET BIN BIN SUM
MODE:C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C D
BIN
:1.0000kHz MANU
V:1.000 V INT.B OFF
RANGE:1 C.V : OFF
DELAY:00000mS AVG : 1
CLR
Press F1.F2
key to the F
F
Press F3 key
to delete the
values input
in error
EXIT
SET BIN BIN SUM
MODE:C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C D
BIN
F:1.0000kHz MANU
:1.000 V INT.B OFF
RANGE:1 C.V : OFF
DELAY:00000mS AVG : 1
CLR
Press F1.F2
key to the V
V
Press F3 key
to delete the
values input
in error
LCR-800 Series User Manual
53
INT.B setting:
l Press F1/F2 key to select the white inverted characters “V”.
l Press F3 or F4 key to select INT or ENT shown as Figure 8-8 below. For
further details, please refer to 4-5-1.
Figure 8-8
INT.B ON OFF setting:
l Press F1/F2 key to select the white inverted characters “ON/OFF” of INT.B.
l Press F3 or F4 key to select ON or OFF shown as Figure 8-9 below.
Figure 8-9
EXIT
SET BIN BIN SUM
MODE:C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C D
BIN
F:1.0000kHz MANU
V:1.000 V OFF
RANGE:1 C.V : OFF
DELAY:00000mS AVG : 1
INT
Press F2 key
to the
INT.B
INT.B
Press F3 .F4
key to select
INT.EXT
EXT
EXIT
SET BIN BIN SUM
MODE:C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C D
BIN
F:1.0000kHz MANU
V:1.000 V INT.B
RANGE:1 C.V : OFF
DELAY:00000mS AVG : 1
ON
Press F2 key
to the OFF
OFF
Press F3 .F4
key to select
ON.OFF
OFF