Tot_BIN setting:
l Press F1/F2 key to select the white inverted characters “Tot_Bin”.
l Key-in needed BIN value through panel keyboard.
l Press ↵ key. Please refer to Figure 8-15 below.
Note: Tot_Bin setting: It can select ascending from 1 to 13 BIN the most.
For example: If BIN1 sorting range is at ±5%, then BIN2 sorting range must
be larger than ±5%, the rest can be done accordingly.
Figure 8-15
C_Nom.Val setting:
l Press F1/F2 key to select the white inverted characters “C_Nom.Val”.
l Key-in Nominal Value through panel keyboard.
l Press ↵ key. Please refer to Figure 8-16 below.
EXIT
BIN1
SORT BY C TOt_Bin:13
:15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max:1.000
D_Min:.0010
CLR
Press F1.F2
key to the
C_Nom.Val
NEXT
BIN
C_Nom.Val
Press F3 key
can delete the
values input
in error
EXIT
BIN1
SORT BY C :13
C_Nom.Val: 15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max:1.000
D_Min:.0010
CLR
Press F2 key
to the
Tot_Bin
NEXT
BIN
Tot_Bin
Press F3 key
can delete the
values input
in error
LCR-800 Series User Manual
59
Figure 8-16
Max setting:
l Press F1/F2 key to select the white inverted characters “Max”.
l Key-in SORT Max value through panel keyboard.
l Press ↵ key. Please refer to Figure 8-17 below.
※After the value is input, it will proceed conversion of +% automatically.
Figure 8-17
“+” setting:
l Press F1/F2 key to select the white inverted characters “+”.
l Key-in the largest percentage of SORT Nominal Value through the panel
keyboard.
l Press ↵ key. Please refer to Figure 8-18 below.
※After the value is input, it will proceed conversion of Max automatically.
EXIT
BIN1
SORT BY C TOt_Bin:13
C_Nom.Val:15.000pF
:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max:1.0000
D_Min:0.00010
CLR
Press F1.F2
key to the
Max
NEXT
BIN
Max
Press F3 key
can delete
the values
input in error
EXIT
BIN1
SORT BY C TOt_Bin:13
C_Nom.Val: 15.000pF
Max:16.500pF :10.0%
Min:13.500pF -:10.0%
D_Max:1.0000
D_Min:0.00010
CLR
Press F1.F2
key to the
+
NEXT
BIN
+
Press F3 key
to delete the
values input
in error