EasyManua.ls Logo

GW Instek PSU 150-10 - Source [:Source]:Current[:Level][:Immediate] Commands [:Amplitude]

GW Instek PSU 150-10
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
95
Source Commands
[:SOURce]:CURRent[:LEVel][:IMMediate]
[:AMPLitude] ......................................................................... 95
[:SOURce]:CURRent[:LEVel]:TRIGgered
[:AMPLitude] ......................................................................... 96
[:SOURce]:CURRent:LIMit:AUTO .................................. 96
[:SOURce]:CURRent:PROTection:DELay ...................... 97
[:SOURce]:CURRent:PROTection[:LEVel] ..................... 97
[:SOURce]:CURRent:PROTection:STATe ...................... 98
[:SOURce]:CURRent:PROTection:TRIPped .................. 98
[:SOURce]:CURRent:SLEWrate:RISing ........................... 99
[:SOURce]:CURRent:SLEWrate:FALLing ....................... 99
[:SOURce]:MODE? ............................................................ 100
[:SOURce]:RESistance[:LEVel][:IMMediate]
[:AMPLitude] ....................................................................... 100
[:SOURce]:VOLTage[:LEVel][:IMMediate]
[:AMPLitude] ....................................................................... 101
[:SOURce]:VOLTage[:LEVel]:TRIGgered
[:AMPLitude] ....................................................................... 101
[:SOURce]:VOLTage:LIMit:AUTO ................................ 102
[:SOURce]:VOLTage:LIMit:LOW .................................. 102
[:SOURce]:VOLTage:PROTection[:LEVel] .................. 103
[:SOURce]:VOLTage:PROTection:TRIPped ................ 103
[:SOURce]:VOLTage:SLEWrate:RISing ........................ 104
[:SOURce]:VOLTage:SLEWrate:FALLing .................... 104
[:SOURce]:CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps. For
externally set current levels (from the analog
control connector) the set current level is returned.
Syntax
[:SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]
{<NR2>(A)|MINimum|MAXimum}
Query Syntax
[:SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]
?
Parameter/Return
<NR2>
0~105% of the rated current output level.

Table of Contents

Other manuals for GW Instek PSU 150-10

Related product manuals