Cause Solution
The instrument does not
receive any buffer or
indicator solution.
Remove the two reagent lines from the Y connector. Only one drop
of reagent comes out per cycle. If no reagent comes out, examine
the pressure plate. Make sure that the pinch block is not
overtightened and the reagent lines are not pinched. Examine the
back of the pinch block for grooves. Replace the pinch block if
necessary.
The chemistry does not
operate correctly.
Mix 1-mL of each reagent and 80-mL of sample. A color change
must show. If no color change shows, replace the reagents.
The instrument does not
operate correctly.
Pinch the sample inlet line to stop the sample flow. The next cycle
reads soft. If it does not read soft, examine the LED if it shows an
orange light. Examine the photocell and clean if necessary. After this
cycle, the instrument reads hard again.
Section 10 Replacement parts and consumables
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage
to the instrument or equipment malfunction. The replacement parts in this section are
approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Refer to the installation kit and maintenance kit documentation for more information.
Replacement parts
Description Item no.
Cam/coupler assembly 5410500
Circuit board assembly 5694800
Colorimeter flow-cell assembly 5516200
Colorimeter drain plug 5103600
Follower block, reagent 4274100
Follower block, sample 4274200
Fuse (T, 1.25 A, 250 V) UL/CSA/CE accepted 5516700
LED 4350800
LED housing, lower 5412700
LED housing, upper 5412800
Maintenance kit 5516500
Installation kit 5516400
Motor assembly 5411900
Photo detector 5411300
Pressure plate 5411800
Stir bar, micro, 3 x 8 mm 5412900