Section 1 Product overview
The Intellical CDC401 series conductivity probes are digital electrodes that measure the electrical
conductivity, salinity, resistivity or total dissolved solids (TDS) of wastewater, drinking water and
general water samples. The probes have a graphite, 4-pole conductivity cell and a temperature
sensor. The standard probes are for laboratory use. The rugged probes are for field use. Refer to
Figure 1.
Figure 1 Probe overview
1 Temperature sensor and conductivity cell 3 Shroud (rugged model)
2 Standard probe 4 Rugged probe
Section 2 Specifications
Specifications are subject to change without notice.
Specifications Details
Probe type Graphite, 4-pole conductivity probe
Conductivity range 0.01 μS/cm to 200.0 mS/cm
Cell constant 0.40 cm
-1
±10%
Conductivity resolution 0.0 to 19.99 μS/cm: 0.01 μS/cm
20.0 to 199.9 μS/cm: 0.1 μS/cm
200 to 1999 μS/cm: 1 μS/cm
2.00 to 19.99 mS/cm: 0.01 mS/cm
20.0 to 200.0 mS/cm: 0.1 mS/cm
Conductivity accuracy ±0.5% of reading
TDS (total dissolved solids) range 0 to 50,000 mg/L as NaCl
TDS resolution 0.0 to 19.99 mg/L: 0.01 mg/L
200 to 1999 mg/L: 1 mg/L
2.00 to 19.99 g/L: 0.01 g/L
20.0 to 50.0 g/L: 0.1 g/L