Krytox oil (item no. 16004801)
Step Description Picture
1 Open the instrument, refer to Open the instrument on page 35.
Remove the optical unit from the mainboard.
2 Insert retainer springs (2x) in the optical cup. Push the springs
in the correct position.
Repairs
38