Step Description Picture
17 The measured value of vial 8 (in absorbance) is shown.
Service inspection finished.
18 If the inspection completes without error, a Certificate of
Service Inspection HDQ303 is automatically stored. Find the
pdf file in the path which is defined in Settings > Program
Settings > General > Certificate path. Print it out and sign.
Refer to Certificate of Service Inspection on page 79.
Subsequently, an instrument log is automatically read out.
Refer to Figure 11 on page 57. Find the txt file in the path
which is defined in Settings > Program Settings > General >
Instrument log path. E.g. Logger_LPG445_18060A000047.txt.
This inspection task is also available as a service video on HSO. HSO full access
requires login credentials. First log into HSO and afterwards click on
https://support.hach.com/app/answers/answer_view/a_id/1024331. Select the desired
service video.
Note: Following hyperlink without being logged will end in a permission denied information
screen.
4.6 Do a service inspection with the test filter set and service mode
Items to collect:
VAA10038.99.V00 Test filter set Service DR300
Certificate of Service Inspection, refer to Certificate of Service Inspection (manually)
on page 81
Step Description Picture
1 Note the part number and serial number from the
DR300 on the Certificate of Service Inspection.
2 Note the set number and expiration date from
VAA10038.99.V00 on the Certificate of Service
Inspection.
Inspection
28

Other manuals for Hach LPV445.99.09110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach LPV445.99.09110 and is the answer not in the manual?

Hach LPV445.99.09110 Specifications

General IconGeneral
ModelLPV445.99.09110
CategoryMeasuring Instruments
ManufacturerHach
Resolution0.01 NTU
Operating Temperature0 to 50°C (32 to 122°F)
Temperature Range0 to 50°C (32 to 122°F)
TypeTurbidity Meter
Measurement Range0 - 1000 NTU
Light SourceLED
MaterialStainless Steel

Related product manuals