EasyManua.ls Logo

Hach ORBISPHERE 510 - Page 78

Hach ORBISPHERE 510
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
77
Communication Menu
Commands
The “Command Output Buffer” is formatted as follows:
The “Command Input Buffer” is located just after the measurement data and is formatted as
follows:
The following commands are available:
Change product
Activate sensor (valid for EC sensors only)
Set channel hold status (valid for TC and LDO sensors only)
Change product command - output
Change product command - input
Name Type Size Offset
Output command toggle (OCT) Output byte 8 bits 0
Output command ID (OCI) Output byte 8 bits 1
Output command data byte 1 (OCD1) Output byte 8 bits 2
Output command data byte 2 (OCD2) Output byte 8 bits 3
Output command data byte 3 (OCD3) Output byte 8 bits 4
Output command data byte 4 (OCD4) Output byte 8 bits 5
Name Type Size Offset
Input command toggle (ICT) Input byte 8 bits 28
Input command status (ICS) Input byte 8 bits 29
Input command data byte 1 (ICD1) Input byte 8 bits 30
Input command data byte 2 (ICD2) Input byte 8 bits 31
Input command data byte 3 (ICD3) Input byte 8 bits 32
Input command data byte 4 (ICD4) Input byte 8 bits 33
Name Value Comment
OCT 1-2
OCI 1 Command ID
OCD1 0
Channel number:
0 = Channel 1
OCD2 0-99 Product number
OCD3 0-1
Erase measurement files:
0 = Never erase the measurement files.
1 = Erase measurement file if necessary (e.g. gas unit changes)
OCD4 Not used
Name Value Comment
ICT 1-2
ICS 0-3
0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no or product number)
3 = Execution failure
ICD1 Not used
ICD2 Not used
ICD3 Not used
ICD4 Not used