78
Communication Menu
Activate sensor command - output
Activate sensor command - input
Set channel hold status- output
Set channel hold status - input
Name Value Comment
OCT 1-2
OCI 2 Command ID
OCD1 0
Channel number:
0 = Channel 1
OCD2 0-1
Sensor activation:
0 = Deactivate the EC sensor
1 = Activate the EC sensor
OCD3 Not used
OCD4 Not used
Name Value Comment
ICT 1-2
ICS 0-3
0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no)
3 = Execution failure
ICD1 Not used
ICD2 Not used
ICD3 Not used
ICD4 Not used
Name Value Comment
OCT 1-2
OCI 3 Command ID
OCD1 0
Channel number:
0 = Channel 1
OCD2 0-1
Status:
0 = Channel enabled
1 = Channel on hold
OCD3 Not used
OCD4 Not used
Name Value Comment
ICT 1-2
ICS 0-3
0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no)
3 = Execution failure
ICD1 Not used
ICD2 Not used
ICD3 Not used
ICD4 Not used