EasyManuals Logo

Hach sensION+ MM150 User Manual

Hach sensION+ MM150
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
23
8. Error messages
Display Caused by Action
Measurement out of range: pH, mV, CE, TDS or
Temperature.
Check sensor.
Low battery charge. Replace the batteries.
E1
Unstable reading during a stability measurement
or during calibration.
100-second timeout.
Check that the sensor is properly immersed in
the sample.
E2
- pH electrode sensitivity out of range. Accepted
values 70-105%
- The difference between the conductivity cell
constants obtained > 30%
Connect a new sensor.
E3
- Asymmetry potential out of range.
Accepted values ± 58 mV.
- During calibration,
he cell constant obtained < 0.05 cm
-1
Connect a new sensor.
E4
Buffer solution not recognised.
Check sensor.
Check buffer solution.
E5
Same buffer solutions.
Check sensor.
Check buffer solution.
E6
The calibration solutions have different
temperatures.
Adjust the calibration solutions to the same
temperature.
ENGLISH
sensION+ MM150 HACH 0710.indd ANG:23sensION+ MM150 HACH 0710.indd ANG:23 7/9/10 13:03:057/9/10 13:03:05

Table of Contents

Other manuals for Hach sensION+ MM150

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach sensION+ MM150 and is the answer not in the manual?

Hach sensION+ MM150 Specifications

General IconGeneral
BrandHach
ModelsensION+ MM150
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals