1 Temperature sensor 2 Reference sensor 3 Measuring sensor
3. The legacy adapter commissioning wizard automatically starts. Follow the instructions that show
on the display. Select the parameter based on the connected sensor.
Option Description
pH Select this parameter if the connected sensor is a pH analogic sensor.
Metal/RedOX/Color Select this parameter if the connected sensor is a Pt-Pt (metallic) analogic sensor or
PTM450/OPT300 sensor.
ISE Select this parameter if the connected sensor is an ion selective sensor.
Refer to the application documentation to enter the related information for the selected
parameter.
Refer to the full user manual for additional information about how to reset the Legacy adapter or
change the sensor name.
Connect the sensor
Use a conical adapter to hold the sensor tightly in the sensor holder.
Connect the sensor to an available sensor port on the rear of the instrument. After the sensor is
connected, make sure that the sensor icon shows in the banner at the top of the display.
N O T I C E
Make sure that the sensor tip is 5 to 10 mm above the top of the magnetic stir bar to prevent any contact with the
bar during operation.
Refer to the illustrated steps that follow.