Figure 2 Instrument components
1 Silicone oil 6 Gelex
®
secondary turbidity standardization kit
2 Oiling cloth 7 StablCal
®
Calibration kit
3 USEPA filter assembly 8 Power supply
4 TL2300 turbidimeter 9 Power cord
5 1-inch sample cells (30 mL) with caps (6x) 10 Dust cover
Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
This instrument is rated for an altitude of 3100 m (10,710 ft) maximum. Use of this instrument at an
altitude higher than 3100 m can slightly increase the potential for the electrical insulation to break
down, which can result in an electric shock hazard. The manufacturer recommends that users with
concerns contact technical support.
Installation guidelines
Install the instrument:
On a level surface
In a clean, dry, well ventilated, temperature controlled location
In a location with minimum vibrations that has no direct exposure to sunlight
In a location where there is sufficient clearance around it to make connections and to do
maintenance tasks
In a location where the power button and power cord are visible and easily accessible
Connect to external devices (optional)
Use the USB ports to connect the instrument to a printer, barcode handset scanner, USB flash drive
or keyboard. Refer to Figure 1 on page 7. The maximum length of a connected USB cable is 3 m
(9.8 ft). As an alternative to the touchscreen, use a keyboard to enter text into text boxes on the
display (e.g., passwords and sample IDs).