20
Start Up
Calibrate curve C-DS1:
7. Select
MEMORY and confirm with ENTER/ON.
8. Lower the probe into a container that contains a homogeneous sample.
9. Select
POINT 1 and confirm with ENTER/ON while stirring the measuring medium with
the probe.
The distance between the probe head and the walls and base of the container must
always be more than 70 mm (2.76 in.).
It takes 5 to 20 seconds to record the calibration point; then the selection menu for
calibration appears. While recording, the display shows the note "Memory".
10. Take the probe out of the container and clean it.
11. Determine the solids content of the sample in the laboratory.
12. Select
*POINT 1 and confirm withENTER/ON.
13. Enter the laboratory value with the navigation keys and confirm with
ENTER/ON.
Navigation key
LEFT/RIGHT: Jump to next/previous decimal place
Navigation key
UP/DOWN: Change number
The laboratory value is saved. The main menu is shown.
Measure:
14. Select
MEASURE, confirm with ENTER/ON.
15. Select
START and confirm with ENTER/ON.
Note: Data is automatically saved every minute, max. 290 measurement points.
Stop measurement:
16. Select
MEASURE, confirm with ENTER/ON.
17. Select
STOP, confirm with ENTER/ON.
Delete the saved measurement points:
18. Select
DATA, confirm with ENTER/ON.
19. Select
DELETE MEP and confirm withENTER/ON.
20. Select YES and confirm withENTER/ON.

Other manuals for Hach TSS Portable

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach TSS Portable and is the answer not in the manual?

Hach TSS Portable Specifications

General IconGeneral
BrandHach
ModelTSS Portable
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals