EasyManua.ls Logo

Hach TU5300 sc - Installation; Installation Guidelines and Overview

Hach TU5300 sc
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 2 Product components
1 TU5300 sc or TU5400 sc 6 Cleaning lid screws and washers for hot water
applications
2 Wall mount bracket (two tubing clips on bracket) 7 Vial replacement tool
3 Tubing clips 8 Flow regulator
4 Tubing clip screws, 2.2 x 6 mm 9 Service bracket
5 Mounting screws, 4 x 16 mm 10 Desiccant cartridge
Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Installation guidelines
N O T I C E
Make sure that there is a floor drain near the instrument. Examine the instrument daily for leaks.
This instrument is rated for an altitude of 3100 m (10,710 ft) maximum. Use of this instrument at an
altitude higher than 3100 m can slightly increase the potential for the electrical insulation to break
down, which can result in an electric shock hazard. The manufacturer recommends that users with
concerns contact technical support.
Installation overview
Figure 3 shows the installation overview with no accessories and the clearances necessary. Refer to
the expanded manual on the manufacturer's website for the system overview with all of the
accessories.