Set the flow rate
1. Measure the flow with the flow regulator fully open. Make sure that the flow is in the middle of the
flow specification. Refer to Specifications on page 3.
2. Slowly close the flow regulator until the flow decreases by 20 to 30%.
Note: The flow regulator causes back pressure in the tubing and decreases the quantity of bubbles that can
form in the vial.
User navigation
Refer to the controller documentation for keypad description and navigation information.
Operation
Refer to the expanded user manual on the manufacturer's website to configure the instrument
settings and to compare process and lab measurements.