Specifications
Specifications are subject to change without notice.
Specification Details
Measurement method Nephelometry with scattered light collected at a 90-degree angle to the incident light
and 360 degrees around the sample vial
Primary compliance
method
DIN EN ISO 7027
Enclosure Material: ASA Luran S 777K / RAL7000, TPE RESIN Elastocon
®
STK40,
Thermoplastic Elastomer TPS-SEBS (60 Shore) and stainless steel
Dimensions (W x D x H) 268 x 249 x 190 mm (10.6 x 9.8 x 7.5 in.)
Weight Instrument with the process head: 2.7 kg (6.0 lb); Instrument with the optional
automatic cleaning module: 5.0 kg (11.0 lb)
Power requirements 12 VDC, 14 VA supplied by the SC controller
Protection class III
Pollution degree 2
Installation category II
Mounting Indoor on a wall
Operating temperature 0 to 50 °C (32 to 122 °F)
Storage temperature –40 to 60 °C (–40 to 140 °F)
Humidity 5 to 95% relative humidity, non-condensing
Sensor cable length 1.6 m (5.25 ft)
1
Laser Class 1 laser product: Contains a non user-serviceable class 1 laser.
Optical light source 850 nm, maximum 0.55 mW
Fittings Sample inlet and outlet: ¼-in. OD tubing
Tubing requirements Polyethylene, polyamide or polyurethane tubing. Calibrated ¼ in. OD,
+0.03 or –0.1 mm (+0.001 or –0.004 in.)
1
Extension cables are available. The maximum sensor cable length is 50 m (164 ft) when no
accessories are connected to the instrument.
English 3