Setup Procedures for Cell Salvage P/N 53063-50, Manual Revision: B
4-10
6. Set up the saline. (Each cycle requires approximately 750 to 1000 ml.)
! Hang the saline wash containers on the IV pole.
! Close the clamps on both of the yellow wash lines before spiking the saline
bags.
! Spike the containers of normal saline, and open the clamps on the lines
Figure 4-14: Hang the saline wash containers on the IV pole
Figure 4-15: Close the ratchet clamps on both spikes before spiking saline
7. Inspect & Finish
! Inspect all parts of the disposable set to be sure that there are no twists,
kinks, or flat spots that would prohibit fluid flow, and that all connections
are secure and all appropriate clamps and covers are closed.