Hanil FLETA 4 - Error Codes

Hanil FLETA 4
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error
code
Error
message
Cause Recommended action
E1 Imbalance
error
- When an imbalance is
detected
- Check weight-balances of samples and then
turn off and on the device for checking.
E2 Over speed
error
- over speed than set speed - Shut off the power supply, and then, turn on
the power switch again to check the device.
- If the error code shows continuously although
you try to operate again, please call please call
Field Service Engineer.
E3 Motor
overheat
- Motor overheat - Shut off the power supply, and then, turn on
the power switch again to check the device.
E4 Motor
operation
error
-Motor operation error - Check that the rotor is rotating through the
RPM measuring window
- Shut off the power supply, and then, turn on
the power switch again to check the device.
E5 Lid
open(Operat
ng)
- Lid open during operation - Shut off the power supply, and contact Hanil
Scientific Inc.
E7 System error - System error - Contact Hanil Scientific Inc.
E11 Motor
temperature
sensor error
Motor temperature sensor
error
- Shut off the power supply, and then, turn on
the power switch again to check the device.
6. Troubleshooting
6.2 Error codes
19