19
If no data was logged for the selected measurement range, the
instrument displays the following messages:
• “NO MANUAL LOGS”
• “NO STABILITY LOGS”
Press GLP/CFM key to enter inside lot information to view recorded
data. Use the keys to toggle between different records.
Use RANGE/u to display GLP data including calibration information, date, time, etc.
Press SETUP/CLR key then GLP/CFM key when deleting records or lots.
Press RCL key to exit the logging type.
Press RCL key to return to the measurement screen.
Delete Logging Type/Lot
Press RCL followed by GLP/CFM key.
Use the keys to select the Manual/Stability records or Interval lots to delete.
Press SETUP/CLR key. The instrument will display “CLEAR MANUAL” for Manual Records, “CLEAR
STAB” for Stability Records.
For Interval lots, the message “CLEAR”, followed by the selected lot will
be displayed with “CFM” tag blinking.
Press the keys to select a different lot. Press GLP/CFM key. The
instrument will display “PLEASE WAIT”.
“CLEAR DONE” is displayed for a few seconds after the selected Interval
lot is deleted.
Delete Records (Manual and Stability log on demand)
To delete individual records (Manual and Stability logs only), enter
Manual (Stability) log by pressing GLP/CFM key when Manual (Stability) is displayed.
Use the keys to sel
ect the record to be deleted and then press SETUP/CLR key.
The instrument will display “CLEAR REC.“ and record number along
with “CFM” tag blinking. Use the keys to select another record
if necessary.
SETUP/INSTALLATION