The “WRONG” message and temperature value are
displayed blinking if the temperature reading is out of
the defined temperature range of the buffer. Calibration
cannot be confirmed in this situation.
Press RANGE to display the
temperature reading on the LCD
during calibration (HI 2211 only).
ONE-POINT CALIBRATION
Proceed as described in “Two-point calibration” section.
Press CAL after the first calibration point
was confirmed.
The instrument will return to measurement
mode and will memorize the one-point
calibration data (new offset).
The instrument will use the slope from the previous calibration. The
default slope will be used if there is no previous calibration.